A-6
Pressure
5 Levels are adjustable, respectively
are:
75±50mmHg (10.0±6.7kPa),
150±125mmHg (20.0±16.7kPa),
300±125mmHg (40.0±16.7kPa),
525±125mmHg (70.0±16.7kPa),
900±180mmHg (120.0±24.0kPa).
Maximum occlusion pressure is about
1080mmHg.
Note 1: The detected pressure of the
infusion system (syringe pump and
syringe) is affected by inner diameter,
piston, material of the syringe and other
factors. Therefore, syringes of different
brand and model may differ in the
detected pressure range.
Note 2: The above declared pressure is
based on Dove and B.Braun Original
Perfusor Syringe (75mmHg is only
available for B.Braun Original Perfusor
Syringe at a rate of ≤100ml/h),
temperature of 20±2
℃
.
525mmHg
KVO Rate
0.1-5.0ml/h
0.5ml/h
Time Near End
Off, 1-30min
when the time is <10min, step for 1min,
and step for 5 min when the time is
≥10min.
3min
Max. Rate Limit
5ml syringe: 0.1-150ml/h
10ml syringe: 0.1-300ml/h
20ml syringe: 0.1-600ml/h
30ml syringe: 0.1-900ml/h
50ml/60ml syringe: 0.1-2000ml/h
5ml syringe: 150ml/h
10ml syringe: 300ml/h
20ml syringe: 600ml/h
30ml syringe: 900ml/h
50ml/60ml syringe:
2000ml/h
Auto-lock Time
Off, 1-5 min, step for 1min
Off
Reminder Time
Off, 1-5 min, step for 1min
2min
Weight Unit
kg, lb
kg
Sound Volume
1-8
4
Brightness
1-8
4
Contrast
150-220
170
Night Mode
On, Off
Off
Nurse call
On, Off
Off
History Record
Can store up to 1500 history records
/
Brand Selection
On, Off
On