A-9
A.6 Infusion Accuracy Curve
The following typical infusion accuracy table expresses performance after infusion
has started and infusion fluctuations occurring within a certain period of time after
normal infusion flow volumes have been reached. The infusion accuracy table is for
reference only; detailed infusion accuracy curve is in accordance with the final
device.
Plotted on the basis of data collected over a two-hour measurement period.
Syringe brand: B.Braun
Original Perfusor Syringe
Syringe specification: 20ml
Sampling quantity of pump: 3
Sampling quantity of syringe: 3
Sampling rate: 1ml/h
Sampling interval:
△
t =0.5 min
Test period: t =120 mins
Infusion rate: Q (m/h)
Syringe brand: B.Braun
Original Perfusor Syringe
Syringe specification: 20ml
Sampling quantity of pump: 3
Sampling quantity of syringe: 3
Sampling rate: 5ml/h
Sampling interval:
△
t =0.5 min
Test period: t =120 mins
Infusion rate: Q (m/h)
NOTE
Infusion accuracy may be influenced by the pump's environment (such as
pressure, temperature, humidity and any infusion consumables used).
Содержание BeneFusion SP3 Vet
Страница 1: ...BeneFusion SP3 Vet Syringe Pump Operator s Manual ...
Страница 2: ......
Страница 20: ...2 5 1 Clip 2 Syringe fixation clamp 3 Slot 4 Spindle clamp 5 Slider 6 Handle 2 2 4 Bottom View 1 Speaker hole ...
Страница 82: ...F 1 F Declaration of Conformity P N 046 013104 00 4 0 ...
Страница 83: ......
Страница 84: ......