6.3.4 Automatic Pressure Release Function (Anti-Bolus)
When occlusion occurs, infusion will stop and the [
Occlusion
] alarm will be
triggered. After the alarm is triggered, the motor is reversed, and the cannula
pressure is then released. This prevents an additional aggressive dose to the animal
after the occlusion is eliminated.
6.4 Key Lock Function
When locked, an
icon in the upper-right corner of the screen emerges. The
following are two ways for automatic lock and manual lock:
Automatic Lock:
1. Select [
Main Menu
]→[
General Options
]→[
Auto-lock Time
].
2. Select [
Auto-lock Time
]: Off, 1-5min. After a specific time is set during the
running state, and if there is no operation or high-level alarm within the set
Auto-lock time, the key board will be auto-locked. [
Off
] indicates switching off
the function.
Manual Lock: In the running interface, under the unlocking condition, press and
hold (>3 seconds)
to lock the keyboard.
Note: If unlocking is needed, press and hold (>3 seconds)
to unlock, it is
automatically locked during the high level alarm.
6.5 Reminder Function
1. Select [
Main Menu
]→[
General Options
]→[
Reminder Time
].
2. Select [
Reminder Time
]: Off, 1-5min. After a specific time is set, the syringe is
loaded. If no operations are performed to the pump within the set time
(including operations on the keyboard, the slider and the pull handle), and the
[
Reminder
] alarm will then alert the user to proceed to the next step. [
Off
]
indicates switching off the function.
6.6 Time Near End
1. Select [
Main Menu
]→[
General Options
]→[
Time Near End
].
2. Select [
Time Near End
]: Off, 1-30min (when <10min, the stair-step is 1min, and
when
≥10min, the stair-step is 5min). After a specific time set, when the
remaining infusion time is close to the [
Time Near End
]
set by the users, [
Time
Near End
] alarm will be triggered. [
Off
] indicates switching off the function.