9-3
NOTE
The lifespan of the battery depends on how frequently it is used and on
how long it has been used, battery capactiy decreases with increase in
charging and discharging times. If the maintenance and storage of the
battery is appropriate, the lifespan of the Lithium battery is no less than
300 times of full charging and discharging. If the use of battery is
improper, its lifespan shall be shortened or in failure status. We
recommend replacing the lithium battery every 3 years.
Please connect to the AC power source if [No Battery] alarm is triggered.
To prevent battery not used for a long time or in No Battery status, if
battery is not charging more than two months after battery is empty,
battery will be in failure status. Do not charge the failure battery, and
replace the failure battery.
If battery will not be used for a long time, we recommend keeping the
battery in a fully charged state and charging the battery every two
months for lifespan guarantee. Please replace the battery if the length of
its lifetime is obviously shortened during optimization.
The length of the battery’s lifetime depends on the device configuration
and operation, for example: Under the condition of the power supply by
the battery, frequent infusion at a high rate will also shorten the length of
the battery’s lifetime.
9.3 Battery Recycling
If there is any obvious damage to the battery or to the battery capacity exhausts, it
should be replaced and recycled appropriately. Please follow the applicable laws on
recycling.
WARNING
The battery must not be disassembled, burned or short-circuited.
Burning, exploding or leaking batteries can cause personal injury.