6-2
6.3 Occlusion Pressure
Occlusion pressure is adjustable, which can meet the requirements for occlusion
pressure of different animals during infusion. Pressure in the infusion tube can be
measured by the built-in pressure sensor, pressure can be calculated by the internal
CPU, which is compared with the preset occlusion alarm threshold. [
Occlusion
]
alarm will be triggered if pressure exceeds the threshold.
6.3.1 Set Occlusion Pressure
1. Select [
Main Menu
]→[
General Options
]→[
Pressure
].
2. Select [
Pressure
]: Occlusion pressure Degree 5, lowest at 75mmHg, and
highest at 900mmHg. Occlusion pressure should be selected according to
actual needs.
CAUTION
If the animal experiences discomfort at a higher occlusion pressure,
monitor the animal's physical conditions under the higher occlusion
pressure closely, and take measures instantly if any abnormal condition
occurs.
6.3.2 Set Pressure Unit
1. Select [
Main Menu
]→[
User Maintenance
]→Input User Maintenance
Password→[
Pressure Unit
].
2. Select [
Pressure Unit
]: The 4 various forms of pressure units, mmHg, kPa, bar
and psi are converted automatically, and can be selected according to actual
needs.
CAUTION
Carefully confirm the edit when changing the current pressure unit.
6.3.3 Dynamic Pressure Scanning (DPS)
During the infusion, the bottom-right corner of the Run screen demonstrates
real-time pressure changes of the animal, in order to find the cannula occlusion at
an earlier time and to prevent the occurrence of further complications.