C-3
[Time Near End]
Low
Infusion remaining time
reaches the setting value
of [Time Near End].
1. The alarm will not be
cancelled automatically
until the infusion is
completed, and then
switch to [VTBI
Complete] or [Syringe
Empty] alarm.
2. Or press
to
cancel alarm.
3. Or press
to
acknowledge the alarm
condition.
[Syringe Near
Empty]
Low
Required time for the
remaining liquid within the
syringe reaches the
setting value of [Time
Near End].
1. The alarm will not be
cancelled automatically
until the infusion is
completed, and then
switch to [VTBI
Complete] or [Syringe
Empty] alarm.
2. Or press
to
cancel alarm.
3. Or press
to
acknowledge the alarm
condition.
[No AC Power]
Low
Power cord disconnects
when the pump has
connected to an AC/DC
power supply.
1. Connect to an AC/DC
power supply to cancel
the alarm.
2. Or press
to
cancel alarm.