A-10
A.7 Trumpet Curve
Flow rate deviation over time (p
△
t)
Syringe brand: B.Braun Original Perfusor Syringe
Syringe specification: 20ml
Sampling quantity of pump: 3
Sampling quantity of syringe: 3
Sampling rate: 1ml/h
Sampling interval:
△
t =0.5 min
Observation windows: p
△
t = 2, 5, 11, 19, 31 mins
Maximum deviation over the course of a full observation window: EPmax (%)
Minimum deviation over the course of a full observation window: EPmin (%)
Average deviation: A (%)
Syringe brand: B.Braun Original Perfusor Syringe
Syringe specification: 20ml
Sampling quantity of pump: 3
Sampling quantity of syringe: 3
Sampling rate: 5ml/h
Sampling interval:
△
t =0.5 min
Observation windows: p
△
t = 2, 5, 11, 19, 31 mins
Maximum deviation over the course of a full observation window: EPmax (%)
Minimum deviation over the course of a full observation window: EPmin (%)
Average deviation: A (%)