6-6
Spline
Spline defines several calculation sections based on the responses of calibration
concentrations. Each section differs in specific parameters. Therefore, the section to
which the current response belongs should be confirmed before Spline calculation.
The parameters of relevant section shall be used to obtain a positive real root with
the dichotomy method.
6.4 QC
Rule
6.4.1 Westgard
Multi-rule
Westgard multi-rule is shown below.
Symbol Explanation
QC
Conclusion
1
2S
One control value
exceeds ±2 standard
deviations.
Warning
1
3S
One control value
exceeds ±3 standard
deviations.
Out-of-control
(random
error,
systematic
error)
2
2S
Two
consecutive
control values for one
level exceed ±2
standard deviations.
Out-of-control
(systematic
error)
R
4S
The
difference
between two
consecutive control
values exceeds 4
standard deviations.
Out-of-control
(random
error)
4
1S
Four
consecutive
control values for one
level exceed ±1
standard deviation.
Out-of-control
(systematic
error)
10
X
Ten
consecutive
control values for one
level lie on one side
of the mean.
Out-of-control
(systematic
error)
Westgard multi-rule QC conclusion flow for single control is shown in Figure 6-3.
Содержание BA-88A
Страница 1: ...BA 88A Semi auto Chemistry Analyzer Operation Manual...
Страница 2: ......
Страница 6: ......
Страница 18: ......
Страница 21: ...III Appendix C External Printer C 1 Appendix D Supplies D 1...
Страница 22: ......
Страница 28: ......
Страница 78: ......
Страница 96: ......
Страница 102: ......
Страница 104: ......
Страница 106: ......
Страница 107: ...P N BA89 20 87038 1 0...