6-5
Multi-point linear calibration
a
b
R
C
−
=
Where,
a
,
b
- calibration parameters
6.3.2 Calculating Concentration of Nonlinearly Calibrated
Sample/control
Logistic-Log 4P
)
)
1
ln(
(
0
b
R
R
K
a
EXP
C
−
−
−
−
=
Where,
0
R
,
K
,
a
,
b
- calibration parameters
Logistic-Log 5P
The positive real root is obtained with the dichotomy method.
Exponential5P
The positive real root is obtained with the dichotomy method.
Polynomial5P
)
)
100
(
)
100
(
)
100
(
exp(
3
0
2
0
0
R
R
d
R
R
c
R
R
b
a
C
−
+
−
+
−
+
=
Where,
0
R
,
a
,
b
,
c
,
d
- calibration parameters
Parabola
The positive real root of the following linear quadratic equation is obtained:
0
2
=
−
+
+
R
c
bC
aC