4-38
4.8.3 Maintenance
In this screen, you can conduct maintenance by selecting
Pump cal
and
Gain
adjust
. The
Password
function is only available to service personnel and authorized
distributors.
Figure 4-26 Maintenance
4.8.3.1 Pump Calibration
Figure 4-27 Pump calibration
Pump Cal
can facilitate the calibration of peristaltic pump. The detailed procedures
are as follows: Prepare 6000µl distilled water, select
1
st
, and then click
Start
. The