6-2
This calibration method requires only one calibrator.
Two-point linear calibration
Calibration formula:
b
aC
R
+
=
.
This calibration method adopts two calibration parameters:
a
and
b
, where,
1
2
1
2
C
C
R
R
a
−
−
=
,
1
1
2
1
2
1
C
C
C
R
R
R
b
)
(
−
−
−
=
.
This calibration method requires two calibrators.
C
1
and
C
2
are respectively the
concentrations of calibrator 1 and calibrator 2.
R
1
and
R
2
are respectively the
responses of calibrator 1 and calibrator 2.
Multi-point linear calibration
Calibration formula:
b
aC
R
+
=
.
This calibration method adopts two calibration parameters:
a
and
b
.
This calibration method requires n (n
≥
3) calibrators.
i
C
is the concentration of
calibrator
i
.
i
R
is the response of calibrator
i
.
a
and
b
can be obtained
through the method of least squares.
n
C
C
n
R
C
R
C
a
n
i
i
n
i
i
n
i
i
n
i
i
n
i
i
i
/
)
(
/
)
)(
(
2
1
1
2
1
1
1
∑
∑
∑
∑
∑
=
=
=
=
=
−
−
=
6.2
Calculating Nonlinear Calibration
Parameters
Logistic-Log 4P
Calibration formula:
)]
ln
(
exp[
1
1
0
C
b
a
K
R
R
+
−
+
+
=
This calibration method adopts four parameters:
0
R
,
K
,
a
and
b
.
This calibration method requires at least four calibrators. The concentration (or
activity) of calibrator 1 is 0, and the corresponding
R
is equal to
R
0
.This
calibration method is applied to the calibration curve that the response becomes
smaller and smaller with the concentration increase. See Figure 6-1.
Содержание BA-88A
Страница 1: ...BA 88A Semi auto Chemistry Analyzer Operation Manual...
Страница 2: ......
Страница 6: ......
Страница 18: ......
Страница 21: ...III Appendix C External Printer C 1 Appendix D Supplies D 1...
Страница 22: ......
Страница 28: ......
Страница 78: ......
Страница 96: ......
Страница 102: ......
Страница 104: ......
Страница 106: ......
Страница 107: ...P N BA89 20 87038 1 0...