Locate the position where the peristaltic pump is installed. Pull out the tubing
that goes through the backboard of the analyzer for 40-50mm until the adapter
is exposed. Pinch the two buckles on the left and right side of the pump shell
and detach the pump head carefully.
3
After detaching the pump head, unplug the tubing (yellow) whose adapters are
closer to the pump head.
Unplug the
tubing
(yellow)
whose
adapters
are closer
to the pump
head.
Pinch the
buckles to
detach the
pump head
Pull the
tubing out
slightly