4-3
Figure 4-3 Basic
The following table explains the parameters on the
Basic
screen.
Parameter Description
Test
The test name should be set and same test names are not allowed.
The length of the test name should not exceed 10 characters.
Otherwise, warning messages will appear.
Reac. Type Analyzing method, including Endpoint, Fixed-time, Kinetics and
Absorbance. The default is Endpoint.
Direction
It refers to the changing direction of the absorbance during the
reaction process, including
Inc
and
Dec
.
If the absorbance increases, select
Inc
; otherwise, select
Dec
.
Pri.
Primary wavelength to be used on the test.
Sec.
Delay
Secondary wavelength to be used on the test. The default is void,
indicating no secondary wavelength applied to the test.
The primary and the secondary wavelength can not be the same.
A period between the end of the aspiration and the start of the test.
Acceptance range is 6-999 sec. The minimum increment or
decrement is 1 sec.