3-2
5
Check if the waster tube at the back of the system is connected to the waste bottle
or other disposal channels. Check if the waste bottle is full. If yes, empty the waste
bottle.
NOTE
Twisted tubing will make the system incapable of aspiration (for system
installed with flow cell). Please check before use, ensuring Ensure the
tubing is not blocked, bent, or twisted.
CAUTION
It will take 15 minutes or more for the lamp and temperature control
system to be stable, after the system is started.
3.2 Timing
Principle
3.2.1 Flow
Cell
Mode
3.2.1.1 Powering On
1
Place the power switch to On.
2
Press the power button of the printer (if internal thermal recorder is
applied, skip this procedure).
3
Place the power switch to On. The system starts initialization which takes
about 2 minutes. During this process, aspirate DI water as directed by the
message on the screen. After initialization and self-check, the main
screen is displayed
3.2.1.2 Setting up the Analyzer
The analyzer will not function properly unless it is properly set up.
If this is the first time the analyzer being used you must complete all the following
settings. In daily use, you should perform the following configuration as needed.
You must enter the hospital information if this is the first time the analyzer being
used. Before requesting test, you must finish the following settings: Basic information,
calibration information and reference range; QC information is recommended.
Click
Setup
Æ
Basic
. Enter the hospital name.
Содержание BA-88A
Страница 1: ...BA 88A Semi auto Chemistry Analyzer Operation Manual...
Страница 2: ......
Страница 6: ......
Страница 18: ......
Страница 21: ...III Appendix C External Printer C 1 Appendix D Supplies D 1...
Страница 22: ......
Страница 28: ......
Страница 78: ......
Страница 96: ......
Страница 102: ......
Страница 104: ......
Страница 106: ......
Страница 107: ...P N BA89 20 87038 1 0...