Environmental and Safety Characteristics
Transport and Storage Temperature:
-40 °C to +70 °C
Transport and Storage Humidity:
5 to 100%, condensing
1
Operating Temperature:
10 °C to 40 °C
Operating Humidity:
10 to 95% RH, non-condensing
(in Airway: 0-100% RH, condensing)
Operating Altitude:
Sea Level to 8,000 feet
Shipping:
ISTA shipping procedure 1A
Shock:
IEC 60068-2-27
peak acceleration: 150 m/s
2
(15.3 g);
duration: 11 ms;
pulse shape: half-sine;
number of shocks: 3 shocks per direction per
axis (18 total).
Vibration:
IEC 60068-2-64
Drop: IEC
60068-2-32
1
After storage in a condensing atmosphere, the unit shall, before use, be kept for more than 24 hr. in an