Gas Module 3 Operating Instructions
2 - 3
Operations
Gas Monitoring with Gas Module 3
WARNING: Connection of the Gas Module 3 exhaust port to the
CAUTION:
O or Anesthetic Agent in the air
surrounding the Gas Module 3 may cause significant
measurement errors.
4.
Check for a clean water trap.
5.
Select
CO
2
or
AUTO
as the
Resp Source
in the
Resp Menu
.
6.
Observe the capnogram on the monitor’s display. On bedside monitor powerup, O
2
,
Agent and N
2
O numbers will display. CO
2
numbers will be displayed when a valid
breath is detected.
NOTE:
The Gas Module 3 must be warmed up a minimum of 45
seconds for ISO accurate CO
2
, O
2
, N
2
O, and agent readings.
7.
If not already set, use the
Display Setup Menu
to select the gas waveforms to be
displayed.
8.
If desired, the gas waveform speed can be changed via the
Monitor Setup
Menu
and the scale can be changed in the
Gas Menu
.
2.1.1
Pre-use Test
Prior to each use, perform the following test with the Gas Module 3 to verify that the gas
analyzer and sample system are functioning properly:
1.
Verify that the appropriate water trap is properly installed and that the appropriate
sampling line is connected.
• DRYLINE
™
Adult/Pediatric water trap used with DRYLINE
™
Adult/Pediatric sampling
line (colorless Luer lock nut)
• DRYLINE
™
Neonatal water trap used with DRYLINE
™
Neonatal sampling line (blue
Luer lock nut)
2.
Verify that the water trap container is less than half full.
3.
Occlude the sampling line and verify that the occlusion alarm functions properly.
4.
Breathe into the sampling line and verify that a CO
2
waveform is correctly displayed on
the monitor.
5.
Sample room air for 30 seconds and verify that the monitor oxygen output is 20.95%
(± sensor inaccuracy).