Electromagnetic Capability
Appendix
5 - 4
Gas Module 3 Operating Instructions
TABLE 5-2
GUIDANCE AND MINDRAY DS USA, INC. DECLARATION - ELECTROMAGNETIC
IMMUNITY
The
Gas Module 3
is intended for use in the electromagnetic environment specified below. The
customer or the user of the
Gas Module 3
should assure that it is used in such an environment.
IMMUNITY
TEST
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact ±8
kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge IEC
61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines IEC
61000-4-11
<5% U
T
(>95%
dip in U
T
) for
0.5 cycle
40% U
T
(60%
dip in U
T
) for
5 cycles
70% U
T
(30%
dip in U
T
) for
25 cycles
<5% U
T
(>95%
dip in U
T
) for
5 sec
<5% U
T
(>95%
dip in U
T
) for
0.5 cycle
40% U
T
(60%
dip in U
T
) for
5 cycles
70% U
T
(30%
dip in U
T
) for
25 cycles
<5% U
T
(>95%
dip in U
T
) for
5 sec
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
Gas
Module 3
requires continued
operation during power mains
interruptions, it is recommended that
the
Gas Module 3
be powered from
an uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
U
T
is the A.C. mains voltage prior to application of the test level.