Gas Module 3 Operating Instructions
3 - 1
3.0
User Maintenance
3.1
Care and Cleaning of Gas Module
WARNING: Do not clean the Gas Module while it is on and/or plugged
1.
The Gas Module enclosure may be cleaned with a mild soap and water solution or
ammoniated window cleaner. Apply cleaning solution to the cloth, not directly onto the
Gas Module. DO NOT apply large amounts of liquid. DO NOT use abrasive cleaning
agents or organic solvents.
CAUTION:
2.
The DRYLINE
™
Water Trap Assembly consists of a filter housing and reservoir that must
be checked and emptied whenever changing patients or if it is more than half full.
WARNING: The contents of the water trap should be handled as a
NOTE:
Replace the complete DRYLINE
™
Water Trap Assembly every
month or more often if indicated on the monitor.
• To remove the DRYLINE
™
Water Trap Assembly from its receptacle, press the lugs on
its sides and pull out. An
Air Leak
message will be displayed. The monitor will
suspend sampling.
• Detach the reservoir from the filter housing by twisting and separating these two parts.
• Empty the reservoir and rinse with water only.
• Tightly re-attach the reservoir to the filter housing.
• Re-install the DRYLINE
™
Water Trap Assembly into the Gas Module, ensuring that it
snaps into place. Check that the
Air Leak
message disappears and monitoring
resumes.