Gas Module 3 Operating Instructions
2 - 7
Operations
Gas Monitoring with Gas Module 3
4.
The message
Feed calibration gas
will be displayed. At this point, attach the
calibration gas canister to the regulator and turn it on. Increasing gas values will appear
in the window as the Gas Module samples the calibration gas.
5.
When calibration is complete, the
Feed calibration gas
message will be removed
from the display and the message
Complete
will be displayed next to each value that
was successfully measured. If at least one gas was successfully measured, the
Accept
menu choice will become available. If the values are acceptable, select
Accept
. To
cancel calibration and re-install the previous calibration values, select
Abort
.
FIGURE 2-10
Gas Calibration Menu
NOTE:
When the “Accept” menu choice is selected, the message
“Disconnect calibration gas.” will be displayed. To avoid
premature emptying of the gas canister, always remove the
regulator at the end of the procedure.
FIGURE 2-8
Gas Calibration Menu
FIGURE 2-9
Gas Calibration Menu