Gas Monitoring with Gas Module 3
Operations
2 - 6
Gas Module 3 Operating Instructions
NOTE:
To avoid premature emptying of the gas canister, always
remove the regulator at the end of the procedure.
NOTE:
If any input data is corrupt or if there are other errors, a
“Calibration Error” message will appear after the “Accept”
button is selected. The Gas Module 3 will not accept span
calibration with errors in any channel.
2.1.2.2
Spectrum OR
™
1.
Select
Calibrate
from the
Gas Menu
. The
Calibration Menu
opens.
FIGURE 2-7
Calibration Menu
2.
Select
Gas Selection
from the
Calibration Menu
and choose the calibration gas
type. Choices are: Mixture, 5% CO
2
, 55% O
2
, 33% N
2
O and 2% Des.
3.
Select
Start
to begin calibration. At the start of the calibration, the message
Zeroing…
will be initially displayed for each of the gas labels as the Gas Module zeros the gas
channels. After successful zeroing, the Gas Module will request the calibration gas as
indicated in the next step.
NOTE:
If the Gas Module cannot zero, a zeroing error will be
displayed and the previous calibration data will be
restored. Repeat the calibration procedure from step 1. If
problems persist, contact Mindray DS Technical Support.