Electromagnetic Capability
Appendix
5 - 6
Gas Module 3 Operating Instructions
TABLE 5-4
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND THE GAS MODULE 3
The
Gas Module 3
is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the
Gas Module 3
can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the
Gas Module 3
as recommended below,
according to the maximum output power of the communications equipment.
RATED MAXIMUM OUTPUT
POWER (P) OF
TRANSMITTER IN WATTS
(W)
SEPARATION DISTANCE (
d
) IN METERS (m)
ACCORDING TO FREQUENCY OF TRANSMITTER
150 kHz to
80 MHz
d
= 1.2 x
80 MHz to
800 MHz
d
= 1.2 x
800 MHz to
2.5 GHz
d
= 2.3 x
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE:
At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
NOTE:
These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
P
P
P