Inrush Function  (Cat. No. 2239-20 only)

Inrush current measurement allows for measuring 
the sharp peak in current upon motor startups.  
Lasting only part of a second, 

MILWAUKEE

's Inrush 

function captures and displays the 

fi

 rst 100 millisec-

onds of current at motor startup.

Power off the system under test.

1. 

Press the jaw opening trigger 

2. 

to open the jaws and clamp 
them onto the conductor 
under test.

 

NOTE: 

Do not clamp 

over 2 or more wires 
at the same time. 
Irregular results 
will occur.
Set the Rotary 

3. 

Dial to

   

or

 

 

position. AC or DC 
mark is displayed.
Press the INRUSH 

4. 

button.
Power on the 

5. 

system under test.
The reading is 

6. 

displayed.

 CAUTION 

Make a measurement 

and press the MIN/MAX button after an 
appropriate range is selected by the auto-
ranging function.

ZERO Function 
(Cat. No. 2239-20 only)
AC Current   / DC Current  :

 

When the LCD doesn’t read 
ZERO while the Jaws are closed, 
press the ZERO Button to in-
dicate ZERO before starting a 
measurement. With the jaws 
closed and without clamping 
them around a conductor, press 
the ZERO key to zero adjust the 
display.
The “ZERO” mark appears on 
the LCD to indicate the ZERO 
Function is activated.

°F / °C  (Cat. No. 2238-20 only)

To switch between Fahrenheit or Celsius, press 
the °F / °C button.

The MIN/MAX recording mode cap-
tures the minimum and maximum 
input values. 
Set the dial to to the desired measure-
ment function. Make a measurement 
and press the MIN/MAX button. The 
meter will capture the MIN and MAX 
readings. Press the MIN/MAX button  
to toggle between the captured MIN, 
MAX and present readings. MAXMIN 

fl

 ashes when the present reading is 

displayed.
To exit and erase stored readings, 
press the MIN/MAX button for two 
seconds or change the dial.
If MIN/MAX is active, all function but-
tons are unavailable except HOLD.

 CAUTION  

The Data Hold readings are 

released when the meter enters Sleep Mode.

Over-

fl

 ow indication

Any time the input exceeds the measuring range 
“OL” or “-OL” is displayed.

 

 Worklight LED ON/OFF

To turn the light on and off, pull the trigger. 

Sleep Mode

Press

Press

0N

0N

D

R

O

T

E

C

T

E

V

A

G

E

O

TL

 

V

O

R

D

E

T

C

T

E

O

E

A

G

LT

0FF

0FF

HOLD Function

Data Hold Function - Freezes the value on the 
display. Press the “HOLD” button to freeze the 
reading. The reading will be held regardless of 
subsequent variation in input. HOLD is displayed 
with the reading. To exit Data Hold mode, press 
the HOLD button again. Hold is not available when 
using the Inrush Function.

MIN/MAX Function 

The clamp meter is automatically powered off in 
about 20 min after the last Rotary Dial or button 
operation. To reset, rotate the Rotary Dial or pull the 
trigger. If the display is still blank when a new Rotary 
Dial setting is selected, charge the battery pack.
To disable the sleep function, turn the tool off. Press 
and hold the HOLD button and turn on the meter. 
After the meter "beeps", release the HOLD button. 
The meter will beep again and the clock symbol  

 will be removed from the display.

The clamp meter does use battery power in sleep 
mode. Be sure to switch the tool to OFF to conserve 
battery power.

 DANGER 

To avoid electrical shock:

Never make measurement on a circuit in 
which voltage over 1000V exists.

FIVE YEAR TOOL  

LIMITED WARRANTY

MILWAUKEE

 Test & Measurement Products (includ-

ing bare tool, li-ion battery pack(s) and battery char-
ger but excluding alkaline batteries) are warranted 
to the original purchaser only to be free from defects 
in material and workmanship.  Subject to certain 
exceptions, 

MILWAUKEE

 will repair or replace any 

part on this product which, after examination, is deter-
mined by 

MILWAUKEE 

to be defective in material or 

workmanship for a period of 

fi

 ve (5) years* after the 

date of purchase.  Return the Test & Measurement 
tool and a copy of proof of purchase to the nearest 
Milwaukee Electric Tool Corporation - factory Service 
Center.  This warranty does not apply to damage that 

MILWAUKEE 

determines to be from repairs made or 

attempted by anyone other than 

MILWAUKEE

 autho-

rized personnel, misuse, alterations, abuse, normal 
wear and tear, lack of maintenance, or accidents.
*The warranty period for the LITHIUM-ION battery 
pack that ships with the Test & Measurement tool is 
two (2) years from the date of purchase.  *Alkaline 
battery that ships with Test & Measurement tool is 
separately warranted by the battery manufacturer. 
*The warranty period for a NON-CONTACT VOLT-
AGE DETECTOR – 2201 20 is one (1) year from 
the date of purchase.
Warranty Registration is not necessary to obtain 
the applicable warranty on 

MILWAUKEE

 product. 

The manufacturing date of the product will be used 
to determine the warranty period if no proof of 
purchase is provided at the time warranty service 
is requested.
ACCEPTANCE OF THE EXCLUSIVE REPAIR 
AND REPLACEMENT REMEDIES DESCRIBED 
HEREIN IS A CONDITION OF THE CONTRACT 
FOR THE PURCHASE OF EVERY 

MILWAUKEE

 

PRODUCT.  IF YOU DO NOT AGREE TO THIS 
CONDITION, YOU SHOULD NOT PURCHASE THE 
PRODUCT.  IN NO EVENT SHALL 

MILWAUKEE

 

BE LIABLE FOR ANY INCIDENTAL, SPECIAL, 
CONSEQUENTIAL OR PUNITIVE DAMAGES, OR 
FOR ANY COSTS, ATTORNEY FEES, EXPENSES, 
LOSSES OR DELAYS ALLEGED TO BE AS A CON-
SEQUENCE OF ANY DAMAGE TO, FAILURE OF, 
OR DEFECT IN ANY PRODUCT INCLUDING, BUT 
NOT LIMITED TO, ANY CLAIMS FOR LOSS OF 
PROFITS.  THIS WARRANTY IS EXCLUSIVE AND 
IN LIEU OF ALL OTHER WARRANTIES OR CON-
DITIONS, WRITTEN OR ORAL, EXPRESSED OR 
IMPLIED.  WITHOUT LIMITING THE GENERALITY 
OF THE FOREGOING, 

MILWAUKEE

 DISCLAIMS 

ANY IMPLIED WARRANTY OF MERCHANTABIL-
ITY OR FITNESS FOR A PARTICULAR USE OR 
PURPOSE, AND ALL OTHER WARRANTIES.
This warranty applies to product sold in the U.S.A., 
Canada and Mexico only.

ACCESSORIES

For a complete listing of accessories refer to your

 

MILWAUKEE 

Electric Tool catalog or go online to 

www.milwaukeetool.com. To obtain a catalog, con-
tact your local distributor or a service center listed 
on the back cover of this operator’s manual.

WARNING 

Always remove battery 

pack before changing or removing acces-
sories. Only use accessories specifically 
recommended for this tool. Others may be 
hazardous. 

Maintaining Tool 

Keep your tool, battery pack and charger in good 
repair by adopting a regular maintenance program. 
After six months to one year, depending on use, 
return the tool, battery pack and charger to a

 

MILWAUKEE

 service facility service.

If the tool does not start or operate at full power 
with a fully charged battery pack, clean the contacts 
on the battery pack. If the tool still does not work 
properly, return the tool, charger and battery pack, 
to a 

MILWAUKEE

 service facility for repairs.

MAINTENANCE

Cleaning

Clean dust and debris from charger and tool vents. 
Keep tool handles clean, dry and free of oil or grease. 
Use only mild soap and a damp cloth to clean the 
tool, battery pack and charger since certain cleaning 
agents and solvents are harmful to plastics and other 
insulated parts. Some of these include gasoline, 
turpentine, lacquer thinner, paint thinner, chlorinated 
cleaning solvents, ammonia and household deter-
gents containing ammonia. Never use 

fl

 ammable or 

combustible solvents around tools.

Repairs

For repairs, return the tool, battery pack and char-
ger to the nearest service center listed on the back 
cover of this operator's manual.

WARNING 

To reduce the risk of per-

sonal injury and damage, never immerse your 
tool, battery pack or charger in liquid or allow 
a liquid to 

fl

 ow inside them.

WARNING 

To reduce the risk of injury, 

always unplug the charger and remove the 
battery pack from the charger or tool before 
performing any maintenance. Never disas-
semble the tool, battery pack or charger. 
Contact a 

MILWAUKEE

 service facility for 

ALL repairs.

Содержание CALIBRATION 2238-20

Страница 1: ...F INJURY USER MUST READ AND UNDERSTAND OPERATOR S MANUAL AFIN DE R DUIRE LE RISQUE DE BLESSURES L UTILISATEUR DOIT LIRE ET BIEN COMPRENDRE LE MANUEL DE L UTILISATEUR PARA REDUCIR EL RIESGO DE LESIONES...

Страница 2: ...nstrument before removing or inserting the battery pack Recharge only with the charger specified by the manufacturer A charger that is suitable for one type of battery pack may create a risk of fire w...

Страница 3: ...ategory III 1000 V Category IV 600 V Pollution Degree 2 EMC EN61326 1 Certifications cULus CE Voltage 12 DC Li Ion MILWAUKEE Battery Pack Cat Nos 48 11 2401 48 11 2402 Battery run time Greater than 12...

Страница 4: ...pe Tem perature Probe to the input terminal The positive side of Probe should be connected to V 3 Place the probe sensor in the desired location 4 The reading is displayed WARNING Never connect the Te...

Страница 5: ...Test Measurement tool and a copy of proof of purchase to the nearest Milwaukee ElectricTool Corporation factory Service Center This warranty does not apply to damage that MILWAUKEE determines to be f...

Страница 6: ...piles si la surface de l instrument est mouill e D brancher tous les cordons d alimentation et les c bles de l objet mis l essai et teindre l instrument avant de retirer ou d ins rer le bloc piles Ne...

Страница 7: ...fils d essai de l instrument pour mesurer le courant Avant l utilisation S assurer que le cadran rotatif est r gl la bonne position que l instrument est r gl au mode de mesure appropri et que la fonct...

Страница 8: ...essai aux deux extr mit s du conducteur mis l essai Si la r sistance mise l essai est de 30 ou moins l avertisseur retentira V T L A O R D O E T C T E G E V T O R G E D E E C T A O T L 12V G T O R E...

Страница 9: ...appuyer sur le bouton MIN MAX minimum maximum apr s avoir s lectionn la plage appropri e l aide de la fonction de plage automatique DANGER Pour viter une d charge lectrique Ne jamais prendre une mesu...

Страница 10: ...el instrumento no funcionan y podr an provocar da os al instrumento o lesiones personales graves Para disminuir el riesgo de lesi n provocada por una descarga y r fagas de arco use equipo de protecci...

Страница 11: ...e voltaje de corr alt voltaje de corr cont 1 200 V de corr alt RMS Rango de corriente de la pinza 720 A de corr alt RMS corr cont 10 s Cat No 2238 20 nicamente 1 200 A de corr alt RMS corr cont 10 s C...

Страница 12: ...C T E O E A G L T PRECAUCI N Cuando la corriente fluye desde el lado de la pantalla hacia la parte inferior del medidor la polaridad es positiva cuando fluye desde la parte inferior hacia el lado de...

Страница 13: ...e Para cambiar entre Fahrenheit y Celsius presione el bot n F C Funci n MIN MAX M NIMO M XIMO El modo de registro MIN MAX m n imo m ximo captura los valores de entrada m nimo y m ximo Configure el sel...

Страница 14: ...ta Mantenga su herramienta en buenas condiciones adoptando un programa de mantenimiento regular Despu s de un a o se recomienda que env e la herramienta a una instalaci n de servicio MILWAUKEE para qu...

Страница 15: ...7 00 AM 6 30 PM Central Time or visit our website at www milwaukeetool com Additionally we have a nationwide network of authorized Distributors ready to assist you with your tool and accessory needs C...

Отзывы: