P93
POS A L/C (B) (V)
Displays the load cell input counts for load cell: B (P 90 = 1), D (P90 = 2), F (P90 = 3), or H (P90 = 4).
Move the test weight to the load cell: A (P90 =1), C (P90 =2), E (P90 = 3), or G (P90 = 4) side of the test
weight bar.
Leave the LCPA-08 SW3 contacts undisturbed.
enter : the displayed value (when stable) to advance to P94.
P94
POS A L/C (A + B) (V)
Displays the summed load cell input counts for:
MSI NUMBER (P90)
LOAD CELLS
1
A and B
2
C and D
3
E and F
4
G and H
For LCPA-08 applications, close the LCPA-08 SW3 load cell contact opened for P92.
Ensure the Test Weight (as applied for P93) is not moved or otherwise disturbed.
enter : the displayed value to advance to P95.
P95
LOAD CELL BALANCE (V)
Displays the count value, which when entered, will balance the load cell signals for the MSI number selected.
For LCPA-08 applications, leave the LCPA-08 SW3 load cell contacts in
the P94 position. Adjust the appropriate LCPA-08 balancing potentiometer until
the P95 display value = 0
±
5. Close all LCPA-08, SW3 contacts.
enter : the displayed value to complete the load cell balance.
P99
RESET
Initiates a complete memory reset.
This feature clears memory locations and restores parameter values to their original factory settings. The
implementation of a Reset requires CompuScale III reprogramming and calibration.
Neither Reset Level affects the Clock / Calender settings. A Partial Reset maintains the previous, P17, P18,
P20, P21, P22, P40, P41, P42 and P43, and load cell balancing settings.
enter:
0 = reset not required (F)
1 = Partial Reset
9 = Master Reset
P100
PARAMETER LIST (V)
Displays parameter names in groups of 10. Pressing "ENTER" advances the display to the next group. This
parameter is not automatically scrolled and therefore must be accessed directly.