P80
V/F CONVERTER OUTPUT (V)
Displays a reference count proportional to the input level from the belt scale. Range = 0 to 131,000.
This feature is normally used to verify the integrity of the belt scale load sensors and cables.
P81
LOAD CELL A (V)
Displays the input from the load cell connected to TB1-1,2 in mV. Range = 0.00 to 50.00.
Access to this parameter is gained when SENSOR TYPE (P21) is set to any value greater than 1. This feature
is normally used to verify the integrity of the load cell after initial installation or subsequent replacement.
P82
LOAD CELL B (V)
Displays the input from the load cell connected to TB1-3,4 in mV. Range = 0.00 to 50.00.
Access to this parameter is gained when SENSOR TYPE (P21) is set to any value greater than 0. This feature
is normally used to verify the integrity of the load cell after initial installation or subsequent replacement.
P83
LOAD CELL C (V)
Displays the input from the load cell connected to TB1-5,6 in mV. Range = 0.00 to 50.00.
Access to this parameter is gained when SENSOR TYPE (P21) is set to 4. This feature is normally used to
verify the integrity of the load cell after initial installation or subsequent replacement.
P84
LOAD CELL D (V)
Displays the input from the load cell connected to TB1-7,8 in mV. Range = 0.00 to 50.00.
Access to this parameter is gained when SENSOR TYPE (P21) is set 4. This feature is normally used to verify
the integrity of the load cell after initial installation or subsequent replacement.
P85
V/F CONVERTER INPUT (V)
Displays the sum of the amplified load sensor (P21) input voltage(s). Range = -1.00 to 4.00.
P88
INITIAL SPAN RESET
Performs a Span Calibration. The results of this calibration establish the new Initial Span count value.
This feature is normally used in response to the [SPAN OUT OF RANGE] message, when the cause of the
deviation is not otherwise correctable. This condition could occur as the result of an intended change in the
Design Parameter values or a mechanical change to the belt or belt scale assembly.
enter:
0 = not required (F)
1 = Initial Span Reset
PL-367
7 – 13