PORT 2 BAUD RATE
The CompuScale III TB2 communication port baud rate is
independant of the PORT 1 baud rate, and is set by SW2
contacts 1 and 2. Set the baud as required by the device
connected. The factory setting is 300 baud.
O = switch contact open
C = switch contact closed
LANGUAGE
Leave SW2 contact 3 (SW2-3) in the open position for English, or closed for alternate language,
CompuScale III displays.
REMOTE COMMUNICATION
Leave SW2 contact 6 (SW2-6) open for normal operation. If a remote mounted CompuScale III is used,
SW2-6 of the remote CompuScale III (only) must be closed.
PORT 2 COMMUNICATION MODE
Leave SW2 contact 7 (SW2-7) open for normal operation, or closed if a serial printer is connected to Port 2.
TEST/ADJUST MODE
Leave SW2 contact 8 (SW2-8) in the open position for normal operation. In the closed position, a test menu is
displayed.
When the keypad test is selected, a graphical representation of the keypad is displayed. As each keypad
button is depressed, the display indicates successful operation.
When the display contrast adjustment is selected, pressing "4" increases, and pressing "8" decreases the
display contrast.
95/11/09
Baud
SW2
Rate
1 2
300
O O
1200
C O
2400
O C
9600
C C
PL-367