The display now shows:
Enter the concentration of the second standard;
e.g.:
[0] [,] [1]
for 0.1.
• Back with the key
[Esc]
.
• EBackout the entry with the key
[F1]
.
Confirm your selection by pressing
[Enter]
.
The display now shows:
Prepare the second standard and press
[Test]
.
The display shows the entered value and the measured
absorbance
value:
Confirm your selection by pressing
[Enter]
.
Note
• To measure further standards, follow the above procedure.
• At least two standards must be measured.
• A maximum of 14 standards (S1 to S14) can be measured.
When the desired number of standards or the maximum num-
ber of 14 standards have been measured, press the key
[Store]
.
Chapter 5.6 | Page 5 of 13
S1 accepted
S2: +_____
ESC F1
S2: 0.10 mg/l
Prepare
Press TEST
S2: 0.10 mg/l
mAbs: 150