Release 08/2014 - Spectroquant
®
Multy
101804
Cell Test
Measuring range: 0.5 –15.0 mg/l Cl
16-mm cell
Chloride
Insert the cell containing
the sample into the cell
compartment. Align the
mark on the cell with
that on the photometer.
Press Test .
Check the pH of the
sample, specified range:
pH 3 – 11.
If required, add dilute
ammonia solution or
nitric acid drop by drop
to adjust the pH.
Insert the blank cell into
the cell compartment.
Align the mark on the
cell with that on the pho-
tometer. Press Zero .
Pipette 10 ml of the
sample into a reaction
cell, close with the
screw cap, and mix.
Add 0.25 ml each of
Cl-1K with pipette,
close with the screw
cap, and mix.
Pipette 10 ml of distilled
water into a second
reaction cell, close with
the screw cap, and mix.
(Blank cell)
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) ready-for-use chloride
standard solution Certipur
®
, Cat.No. 119897, concentra-
tion 1000 mg/l Cl
-
, can be used after diluting accordingly.
Select method 1 2 2 .
Reaction time:
10 minutes
Press Enter to start
the countdown.