![Millipore Spectroquant Multy Скачать руководство пользователя страница 273](http://html1.mh-extra.com/html/millipore/spectroquant-multy/spectroquant-multy_operating-manual_1796903273.webp)
The display now shows:
Enter the pH within the range 0 and 12 and confirm by
pressing
[Enter]
.
The display now shows the Langelier saturation index:
Pressing the
[Enter]
key starts the entry mode anew (entry of
the temperature result).
Pressing the
[Esc]
key takes the instrument back to the mode
menu.
Note:
If a result is entered that is beyond the defined range of
entries, an additional message appears in the display, e.g.
Value too high.
Value too low.
Acknowledge this message by pressing
[Enter]
and enter a
value that is within the defined range.
Chapter 5.8 | Page 3 of 3
<Langelier>
pH value
0 <= pH <= 12
+ _ _ _ _
<Langelier>
Langelier
saturation index:
-0.25
ESC
<Langelier>
Calcium hardn.
50<=CH<=1000
CH<=1000 mg/l CaCO3 !
<Langelier>
Calcium hardn.
50<=CH<=1000
CH>=50 mg/l CaCO3 !