![Millipore Spectroquant Multy Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/millipore/spectroquant-multy/spectroquant-multy_operating-manual_1796903051.webp)
Add to each cell 1 level
blue microspoon of
Al-1K, close with the
screw cap.
Measuring range: 0.05 – 0.50 mg/l Al
16-mm cell
Add to each test tube
0.25 ml of Al-2K with
pipette, close with the
screw cap, and mix.
100594
Cell Test
Aluminium
Pipette 6.0 ml of the
sample into a reaction
cell, close with the
screw cap, and mix.
Pipette 6.0 ml of distilled
water into a second
reaction cell, close with
the screw cap, and mix.
(Blank cell)
Shake both cells vigor-
ously to dissolve the
solid substance.
Select method 2 1 .
Reaction time:
5 minutes
Press Enter to start
the countdown.
Insert the cell containing
the sample into the cell
compartment. Align the
mark on the cell with
that on the photometer.
Press Test .
Insert the blank cell into
the cell compartment.
Align the mark on the
cell with that on the pho-
tometer. Press Zero .
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) ready-for-use aluminium
standard solution Certipur
®
, Cat.No. 119770, concentra-
tion 1000 mg/l Al can be used after diluting accordingly.
Check the pH of the
sample, specified range:
pH 3 – 10.
If required, add dilute
sodium hydroxide
solution or sulfuric acid
drop by drop to adjust
the pH.
Release 08/2014 - Spectroquant
®
Multy