Pipette 5.0 ml of
distilled water (Water
for analysis EMSURE
®
,
Cat.No. 116754, is
recommended) + 5.0 ml
of the sample into a
24-mm cell (do not add
any reagents!), close
with the screw cap, and
mix.
(Blank cell)
Add 1 tablet Cyanuric
Acid, crush with stirring
rod, and close with the
screw cap.
Swirl the cell to dissolve
the solid substance.
Add 5.0 ml of distilled
water (Water for analy-
sis EMSURE
®
, Cat.No.
116754, is recommen-
ded) with pipette, close
with the screw cap, and
mix.
Pipette 5.0 ml of the
sample into a 24-mm
cell.
Release 08/2014 - Spectroquant
®
Multy
Measuring range: 2 – 160 mg/l cyanuric acid 24-mm cell
Cyanuric Acid
119250
Test
Filter turbid samples.
Select method 2 0 0 .
Insert the blank cell into
the cell compartment.
Align the mark on the
cell with that on the pho-
tometer. Press Zero .
Insert the cell containing
the sample into the cell
compartment. Align the
mark on the cell with
that on the photometer.
Press Test .
Quality assurance:
To check the measurement system (test reagents, meas-
urement device, and handling) a cyanuric acid standard
solution must be prepared from Cyanuric acid, Cat.No.
820358 (see section “Standard solutions”).