Performing an Analysis
ASAP 2460 Operator’s Manual
2-42
246-42800-01 - Aug 2013
Krypton Analysis
Use the following guidelines for performing krypton analyses:
• All ports must be idle in order to start an analysis.
• All analyses must use krypton gas.
• From one to five analyses can be started at the same time. The last port is used to store Krypton for
dosing (Port 2, 4, or 6 depending on your system configuration).
• An empty sample tube must be installed in the last port (2, 4, or 6 depending on your system
configuration).
• The sample file used for an analysis must specify the Dosing Method:
From last port
in the
Adsorptive Properties.
• Samples cannot be removed from or added to ports until the full set of analyses has
completed
The data collection is done sequentially — one analysis starts and completes before the next is started.
1.
Install an empty sample tube on the port to be used to store krypton (the last port: 2, 4, or 6) and
place a dewar cover over the sample tube stem and the p
o
) tube stem.
Krypton analyses are available only if the Krypton option is installed on the analyzer.
The dewar below the port used to store and purify krypton must be at least 50% full
of the analysis bath liquid and an empty sample tube must be installed on the port
before starting a krypton analysis.
Содержание ASAP 2460
Страница 4: ......
Страница 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Страница 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Страница 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Страница 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Страница 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Страница 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Страница 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Страница 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Страница 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Страница 234: ...Report Examples 2460 Operator s Manual 5 32 246 42800 01 Aug 2013 BJH Adsorption Cumulative Pore Volume...
Страница 235: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 33 BJH Desorption Cumulative Pore Volume...
Страница 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Страница 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Страница 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Страница 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Страница 278: ......
Страница 280: ......
Страница 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Страница 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Страница 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Страница 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Страница 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Страница 415: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 422: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...