4017-
Damage to the instrument will result if the sample has not been manually evac-
uated. Have you evacuated the sample?
Cause:
You did not select
Backfill sample at start of analysis
on the Sample
Backfill Options dialog box. The sample tube is normally at atmo-
spheric pressure when an analysis is started, and it must be backfilled
before the analysis begins to prevent sample material from being
drawn into the manifold.
Action:
If you have manually evacuated the sample tube, select
Yes
. If you
have not, select
No
and then either perform a manual evacuation or go
to the Sample Backfill Options dialog box and select
Backfill sample
at start of analysis
.
4020-
Disabling this option may damage the instrument. Are you sure that the sample
should not be backfilled?
Cause:
You did not select
Backfill sample at start of analysis
on the Sample
Backfill Options dialog. The sample tube is normally at atmospheric
pressure when an analysis is started; it must be backfilled before the
analysis begins to prevent sample material from being drawn into the
manifold.
Action:
If you want to manually evacuate the sample prior to the start of the
analysis, select
Yes
. Otherwise, select
No
, go to the Sample Backfill
Options dialog, and select
Backfill sample
at start of analysis.
4021-
The entered Po value (Po and Temperature Options of the Analysis Conditions)
is outside the range of the pressures listed in the Psat vs. Temperature Table (Adsorp-
tive Properties).
Cause:
The temperature that corresponds to the entered Po cannot be found
because the entered Po is outside of the range of values in the Psat vs. T
table.
Action A:
Enter a Po that is within the range of the table.
Action B:
Use a different Psat vs. T table.
Содержание ASAP 2460
Страница 4: ......
Страница 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Страница 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Страница 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Страница 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Страница 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Страница 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Страница 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Страница 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Страница 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Страница 234: ...Report Examples 2460 Operator s Manual 5 32 246 42800 01 Aug 2013 BJH Adsorption Cumulative Pore Volume...
Страница 235: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 33 BJH Desorption Cumulative Pore Volume...
Страница 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Страница 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Страница 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Страница 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Страница 278: ......
Страница 280: ......
Страница 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Страница 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Страница 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Страница 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Страница 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Страница 415: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 422: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...