Start Analysis
ASAP 2460 Operator’s Manual
4-4
246-42800-01 - Aug 2013
Start Analysis
This option allows you to start an analysis using nitrogen or a similar gas. Use the following guidelines
for performing standard analyses:
• Standard analyses cannot be performed if a high throughput or krypton analysis is in progress.
• One analysis can be started at a time.
• A sample can be added to any idle port and an analysis started without disturbing the analyses being
performed on other ports.
• Samples can be removed from any of the six ports without disturbing the analyses being performed
on other ports.
• The sample dosing method should be
Normal
.
• All analyses must use the same gas.
• If saturation pressure is being measured, all analyses must use the same Psat gas.
• A manual evacuation can be performed on any idle port while analyses are in progress.
Report after analysis
button (
.REP
(Report system) - saves the report in a format that can be opened
within any MicroActive program.
.TXT
(ASCII text) - saves the report as a text file.
.XLS
(Spreadsheet file) - saves the report in a format that can be
opened within a spreadsheet program.
Start
button
Click to start an analysis or calibration procedure.
Field or Button
Description (continued)
Содержание ASAP 2460
Страница 4: ......
Страница 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Страница 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Страница 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Страница 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Страница 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Страница 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Страница 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Страница 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Страница 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Страница 234: ...Report Examples 2460 Operator s Manual 5 32 246 42800 01 Aug 2013 BJH Adsorption Cumulative Pore Volume...
Страница 235: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 33 BJH Desorption Cumulative Pore Volume...
Страница 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Страница 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Страница 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Страница 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Страница 278: ......
Страница 280: ......
Страница 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Страница 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Страница 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Страница 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Страница 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Страница 415: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 422: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...