Appendix B
ASAP 2460 Operator’s Manual
B-30
246-42800-01 - Dec 2013
4069-
Dosing method choice is invalid. The Analysis Conditions choice of Absolute
pressure dosing requires that the Adsorptive Properties (Dosing Method) is set to
Normal.
Cause:
There is an incompatibility between the analysis conditions choice of
Absolute pressure dosing
and the adsorptive properties
dosing
method.
Action:
Edit one of the choices.
4070-
Unable to load deconvolution model (name).
Cause:
For some reason, the list of available models was corrupted, therefore,
the model selected could not be loaded for the deconvolution.
Action:
Exit the program and reinstall the software, then try again.
4071-
The selected pressures points do not form a valid set for deconvolution.
Cause:
The data points selected for analysis do not contain enough informa-
tion to allow a DFT data reduction.
Action A:
Edit data points in the table of the Collected Data dialog, or select
another sample file. At least two points with strictly increasing pres-
sures and volumes adsorbed are required for a DFT Plus data reduc-
tion.
Action B:
Edit the pressure range on the DFT report options dialog.
4072-
The range of pressures selected is too small to deconvolute using this model.
Cause:
A null result was found using the selected model.
Action:
Edit data points in the table of the Collected Data dialog, or select
another sample file. At least two points with strictly increasing pres-
sures and volumes adsorbed are required for a DFT data reduction.
Содержание ASAP 2460
Страница 4: ......
Страница 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Страница 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Страница 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Страница 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Страница 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Страница 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Страница 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Страница 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Страница 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Страница 234: ...Report Examples 2460 Operator s Manual 5 32 246 42800 01 Aug 2013 BJH Adsorption Cumulative Pore Volume...
Страница 235: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 33 BJH Desorption Cumulative Pore Volume...
Страница 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Страница 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Страница 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Страница 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Страница 278: ......
Страница 280: ......
Страница 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Страница 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Страница 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Страница 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Страница 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Страница 415: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 422: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...