3.4.5 To read out ELC serial number and software version
These data may be accessed via display or PROFIBUS and may be of interest e.g. for service
and in case of a defect.
1. Display
Drücken < 5sek
Drücken < 5sek
Shortly press the SET-button to reach the next menu
item.
Lamp operation data see paragraph 4.4.
Serial number and currently installed software version
in decimal values.
2. PROFIBUS-input-data (ELC
→
Master)
Byte-
No.
0 1 2 3
4
5
6
7
8
9
10 11 12
13
14
15
Name [ELC
Serial
Number]
[Software
Version]
[ELC
Nominal
Power]
[Error
Register]
[Warning
Register]
[Lamp
Status]
[Actual
Power]
[Actual
Voltage]
[Actual
Current]
[Reserved]
↓
Byte 2
Byte 1
Byte 0
7
6
5
4
3
2
1
0
7
6
5
4
3
2
1
0
7
6
5
4
3
2
1
0
[ELC Serial Number
is indicated by the first three registers. The exact division of the bits into
year, month and consecutive number is defined by the brackets.
Example: 01F305
16
stands for: consecutive number = 62, month = 6 and year = 5.
[Software Version]
contains the currently installed software version. The high nibble
represents the main version (e.g. 10), the low-nibble the secondary version (e.g. 2).
Accordingly the example Version 10.2 is written as A2
16
.
month
year
consecutive number
ELC ready
Bus connected 19h
Serial Number
Software Version
Lamp off