5.11 Description of the remote control commands
774 Oven Sample Processor, Instructions for use
129
5.11.6.7
A
ssembly
.Sc
an $G,
$S
A
ssembly
.Sc
an
.A
ddress
Rm
, RS
A
ssembly
.Sc
an
.P
attern
with Rm (parallel/Remote):
8 x 1, 0 or
∗
(bin)
ready1
, end1, Cond ok.
Cond 737, no error
with RS (serial/RS232):
14 ASCII characters
Scanning the interfaces.
.Address
Selection of interface (Remote / RS232)
.Pattern
Signal or character string
This function is not applicable for process control via RS232 interface. See
&Info.ActualInfo.Inputs and …Outputs (5.11.3.14 and 5.11.3.15) instead.
5.11.6.8
A
ssembly
.C
trl $G
A
ssembly
.C
trl
.A
ddress
Rm
, RS
A
ssembly
.C
trl
.P
attern
with Rm (parallel/Remote):
14 x 1, 0 or
∗
(bin),
START device1, START dos1,
STOP device1, ENTER,
INIT
with RS (serial/RS232):
14 ASCII characters
Signal or data transmission via interfaces.
.Address
Interface selection (Remote/RS232)
.Pattern
Signal or character string
With the Remote interface predefined bit patterns may be applied using the
short names listed above (see page 99ff).
5.11.6.9
A
ssembly
.De
f $G
A
ssembly
.De
f
.O
bject
DOSRATE
FILLRATE,
LIFTRATE
SHIFTRATE,
DRIVE.PORT
A
ssembly
.De
f
.A
ddress
depending on .object
A
ssembly
.De
f
.V
alue
depending on .object
Definition of various device settings. Depending on the selected item in
…DEF.Object different parameters and ranges may be entered.
Def.Object
Def.Address
Def.Value
DOSRATE
1
…12 0.01…
160
ml/min
FILLRATE
1
…12 0.01…
160
ml/min
LIFTRATE
1
…2 3…
25
mm/s
SHIFTRATE
auto.
, +, –
3…
20
ang. degrees/s
DRIVE.PORT [
1
…12].[
1
…4]
dos.
, fill, rinse, prep., drain
&Assembly.Def;$G triggers this function.
5.11.6.10
A
ssembly
.W
ait
$G, $S, $H, $C
A
ssembly
.W
ait
.T
ime 0...
1
...9999 s
Waiting time.
&Assembly.Wait;$G triggers this function.
5.11.6.11
A
ssembly
.E
nd $G
$G triggers a RESET (same as the <RACK> key).
Содержание 774
Страница 2: ......
Страница 6: ......
Страница 106: ...5 Detailed description 774 Oven Sample Processor Instructions for use 100 ...