5 Detailed description
774 Oven Sample Processor, Instructions for use
120
5.11.2.5
C
onfig
.Ra
ckDef
.Spez
Beak
.1.P
os
0
...number of rack positions
etc.
until
.8
Rack positions of special beakers 1 to 8 (position 0 = not defined).
5.11.2.6
C
onfig
.P
osTab
.TabI
dx
0
...31
Index of position tables..
5.11.2.7
C
onfig
.P
osTab
.N
ame
8 ASCII characters
C
onfig
.P
osTab
.R
1Num
2...(R2Num – 2)
C
onfig
.P
osTab
.R2
Num
(R1Num + 2)...(R3Num – 2)
C
onfig
.P
osTab
.R3
Num
(R2Num + 2)...200
C
onfig
.P
osTab
.R1O
ff
0
...3599
C
onfig
.P
osTab
.R2O
ff
0
...3599
C
onfig
.P
osTab
.Nu
m
1
...200
Definitions of position tables. Depending on the table index the correspond-
ing one of the 16 possible data sets is overlayed.
Name:
Identification of the rack type. Will be available as a selec-
tor under &Config.RackDef.Type (see 5.11.2.4).
R1Num:
Highest Beaker position in row 1
R2Num:
Highest Beaker position in row 2
R3Num:
Highest Beaker position in row 3
R1Off Offset
in
1
/
10
-angular degrees for the beaker positions in
row 1 (for the beaker test)
R2Off Offset
in
1
/
10
-angular degrees for the beaker positions in
row 2 (for the beaker test)
Num:
Number of rack positions
5.11.2.8
C
onfig
.P
osTab
.1.V
alue
0
...3599
bis
.200.V
alue
Angular offset for the particular rack position in tenth of angular degrees
(
1
/
10
-degrees).
Definition of new rack types:
The definition of rack types is only possible via RS232 interface.
•
Set table index (see 5.11.2.6).
•
Enter name of rack type (&Config.PosTab.Name, see 5.11.2.7).
•
Enter highest beaker position in row 1 to 3 (&Config.PosTab.R1–3Num,
see 5.11.2.7).
•
Enter the offset angle between beaker position 1 in row 1 and beaker
position (R1num+1) in row 2, respectively and tower 1
(&Config.PosTab.R1–2Off, see 5.11.2.7).
•
Define number of rack positions (&Config.Pos.Tab.Num, see 5.11.2.7).
•
Enter the particular angular offset for each rack position
(Irregular layouts of rack positions are possible)
5.11.2.9
C
onfig
.W
etPart
.W
etPartNo
1
...12
Identification of dosing unit.
5.11.2.10
C
onfig
.W
etPart
.1.M
axRate 0.01...
160
ml/min
C
onfig
.W
etPart
.1.L
ength 0...
1000
...30000 mm
C
onfig
.W
etPart
.1.D
iameter 0.1...
2
...20 mm
until
.4.D
iameter
(Default value .
2
.
L
ength :
250
mm)
Содержание 774