5 Detailed description
774 Oven Sample Processor, Instructions for use
54
Key
Normal state and editing
•
DEF commands are valid for manual operation as well as for the
programmed processing of a method.
•
The individual DEF commands are listed below.
0
DEF
DOSRATE
Change dosing rate
•
The dosing rate may be individually set for every dosing drive (Do-
simat or Dosino).
•
Syntax: DOSRATE [Dosing unit] [Dosing rate]
0
DEF
FILLRATE
Change filling rate
•
The filling rate may be set individually for every dosing drive (Dosi-
mat oder Dosino).
•
Syntax: FILLRATE [Dosing unit] [Filling speed]
0
DEF
LIFTRATE
Change lift speed
•
The lift speed can be set at will.
•
Syntax: LIFTRATE [Tower] [Lift speed]