5 Detailed description
774 Oven Sample Processor, Instructions for use
72
'EoD' (End of Determination) pulse at the remote link. The SCAN timeout de-
fines the maximum permitted duration of a determination in this case.
The timeout time starts directly on activation of the SCAN command. After the
timeout time has expired, an error message is displayed and the sequence of
the method is aborted (HOLD status) or the system continues with the next
command of the sequence, depending on the setting above.
Parameter
>gas flow
Submenu for gas flow settings
Open the submenu with <ENTER>
>gas flow
unit gas flow:
mL/min
Selection of the unit for display of the
gas flow rate
mL/min
, L/h
>gas flow
min. flow
0.0 mL/min
Lower warning limit of flow rate
0,0
…999 mL/min
>gas flow
max. flow
900 mL/min
Upper warning limit of the flow
rate
0,0…
900
…999 mL/min
>gas flow
gas type
air
Selection of the carrier gas
air
,N2, other
>gas flow
gas flow factor
1
Correction factor for gas flow meas-
urement in the case of 'other' gases
0,001…
1
…9,9
Carrier gas:
Argon
Methane
CO
2
Natural gas
Helium
N
2
O
Oxygen
Propane
Factor:
1.456
0.717
0.738
0.681
1.456
0.666
0.992
0.357