5.8 Remote Interface
774 Oven Sample Processor, Instructions for use
99
CTL-command
Parameter Bit
Pattern
Function
INIT
00000000000000
initializes the remote interface
START device1
*************1
starts device1 (for ex., Titrino, Coulometer...)
START dos1
*******1******
starts Dosimat on device1 (Titrino via "activate",
6.2141.040 cable required)
STOP device1
************1*
stops device1 (for ex., Titrino, Coulometer...)
START 737
*********1****
starts Coulometer 737 (6.2141.000 cable required)
ENTER
*********1111*
simulates the <ENTER> key
of
Titrinos or a
756 Coulometer
With the START and STOP commands a short pulse (200 ms) is emitted.
Manual stop options
Parameter Bit
Pattern
Function
–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
STOP device1
************1*
stops device1 (for ex., Titrino, Coulometer...)
With the STOP command a short pulse (200 ms) is emitted.
SCN Command
Parameter Bit
Pattern
Function
–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
Ready1
*******1
queries "ready" state of device1 (Titrinos. …)
End1
****1***
expects the end pulse of device1 (EOD)
Cond ok
******1*
queries "Cond" state of device1 (Titrinos. …)
Cond 737
1****1**
queries "Cond" state of the 737 Coulometer
no error
**1*****
queries "no error" state of device1 (Titrinos. …)