
6.6 Validation / GLP
774 Oven Sample Processor, Instructions for use
153
You can use a water standard on an anisole or xylene basis (1.00 mg resp.
0.10 mg water/mL) for validation of a 774 Oven Sample Processor at lower
temperatures (<150 °C). The oven temperature must be lower than the boil-
ing point of the solvent used.
Recommended settings:
Oven temperature: 120 °C
Gas flow:
Approx. 60 mL/min
Sample size:
Approx. 1 g
Содержание 774
Страница 2: ......
Страница 6: ......
Страница 106: ...5 Detailed description 774 Oven Sample Processor Instructions for use 100 ...