E-BAM-9805 Manual Rev B.docx
Page 9
2 E-BAM INSTALLATION AND COMMISSIONING
This section describes assembly, start-up, and filter tape installation for the E-BAM unit.
2.1 Unpacking
Any damage incurred to the equipment during shipping are the responsibility of the carrier. If any
damage to the shipment is noticed before unpacking,
a claim must be filed with the commercial
carrier immediately
. Follow any special unpacking instructions provided by the carrier as all items
are then carefully removed from the containers. Inspect each component thoroughly. It is
recommended to document and photograph all damaged packages and items before, during, and
after unpacking them. Contact Met One Instruments, Inc. (see section 1.2 of this manual) to arrange
for any replacement items needed.
Unpack the mass monitor and accessories and compare them to the packing list to make sure all
items are present.
The E-BAM is shipped with the nozzle shipping device installed under the nozzle which prevents
the moving parts of the vane assembly from being damaged in transit. The device must be replaced
anytime the mass monitor is being transported in order to avoid damaging the vane control
mechanism. Do not ship or transport the E-BAM with filter tape installed.
Please keep all special shipping items (box, foam packing material, etc.) used to ship the E-BAM.
They should be re-used if the monitor is transported (changing site locations, returning to the
factory, etc.). Met One Instruments, Inc. is not responsible for damage to the mass monitor if
shipped in non-original packaging or without the shipping shim in place. Contact Met One
Instruments, Inc. (see section 1.2 of this manual) for replacement packing materials, if necessary.
2.2 Accessories
Every E-BAM requires the use of ancillary components. The items supplied with a standard E-BAM
PM
10
monitoring system are shown in Table 2-1.
Standard Components
Part No.
Mounting Tripod
EX-905
PM
10
Size Selective Inlet
BX-802
Inlet Tube (10 inches)
81183
Ambient Combination Sensor (AT, BP, RH)
597A w/ 82170 Cable Assembly
Filter Tape
460180
Accessory Kit*
83230
*See section 7 for a complete listing of all items contained in the Accessory Kit
Table 2-1 Standard E-BAM Accessories