3.6.9 Station ID
This screen provides the option to enter a three-
digit location ID for the E-BAM. Press the green
bordered value box and the numerical entry
keypad (Figure 3-5) will be displayed. Enter the
new location ID value.
Figure 3-32 The Station ID Screen
3.6.10 MET Average
The default averaging period for the E-BAM is 60 minutes. If an alternate averaging period is
needed, it may be selected from this screen. Press the green bordered value box and the selection
screen shown on the right of Figure 3-33 will be displayed. Not all choices are able to be presented
at the same time. Use the up and down arrow keys under the displayed options to show additional
selections Tap the required period length and then press the OK button to set it. The available MET
average periods are 1, 5, 10, 15, 30, and 60 minutes.
Figure 3-33 The MET Average Screens
3.6.10.1
Data Average Capacity
The E-BAM internal data logger can store 22,528 records. With a fixed number of records, the time
until the internal logger is full will vary depending on the averaging period selected. The shorter the
time interval, the faster the logger will fill up. Table 3-3 shows the approximate length of time in
days, months, and years before the logger reaches full capacity. Once the circular memory is full,
the E-BAM will begin overwriting the oldest data with the most recently collected data points.
Time Period in
Minutes
Approximate
Capacity in Days
Approximate
Capacity in Months
Approximate
Capacity in Years
1
15.6
0.5
0.0
5
78.2
2.6
0.2
10
156.4
5.2
0.4
15
234.7
7.8
0.6
30
469.3
15.6
1.3
60
938.7
31.3
2.6
Table 3-3 Approximate Data File Capacity in Time