5.12 E-BAM Hardware Failure Screen
This screen is presented to the user when a
hardware failure occurs with the type of alarm
(Tape Break, Sensor Range, Flow Failure, etc.)
indicated in red text. Pressing the grey ALARMS
button will present the View Alarms screen. See
section 0 for more information about viewing
alarms.
Figure 5-14 Sample Alarm Screen
5.12.1 Sensor Out of Range Event
The sensor outside range alarm event occurs when one of the E-BAM sensors is reading outside of
its designated limit. Each sensor, along with its upper and lower range limits, is shown in Table 5-5.
The Stop column indicates whether or not the error will cause the E-BAM to stop sampling.
Sensor
Minimum Maximum
Stop
AT
-50
70
Yes
RH
0
100
No
BP
375
825
Yes
UPPER
228
818
Yes
LOWER
228
818
Yes
FT
-40
60
Yes
FRH
0
98.9
No
BOX
-40
60
No
BETA
500
20000
Yes
Table 5-5 Sensor Ranges