E-BAM-9805 Manual Rev B.docx
Page 58
6.4 Downloading Data Using HyperTerminal or other Terminal Programs
The E-BAM is able to communicate via serial commands using RS-232, RS-485, or a USB Serial.
The monitor only allows one of these connections at a time. These connections allow users to both
download data and operate the instrument
The default RS-232 settings are as follows:
Baud Rate: 115200
Data Bits:
8
Parity: None
Stop Bits:
1
Flow Control: None
6.4.1 User Communication
In the User Communication (or Terminal) mode, commands are issued by entering a letter or
number command and then pressing the Enter, <cr>, key. The <Esc> key is not used when sending
commands in this mode.
Press the Enter key three times to enter User Communication mode. An asterisk character appears
confirming the mode has been entered and that the instrument is ready for a new command.
Commands will be echoed back from the instrument and an asterisk will appear after a command
has completed.
COMMAND
DESCRIPTION
1
Request settings report
2
Request All data report
3
Request New data report
4
Request Last data report
7
Request alarm event report
C
Clear data log file
– $
D
Get/Set date part of the real time clock
– $
H
Help menu
Q
Quit and exit terminal mode
T
Get/Set time part of the real time clock
CA
Clear alarm log file
– $
DT
Get/Set the date and time of the real time clock
ID
Get/Set location ID or address
QH
Report current readings header
RQ
Report current readings without header
RV
Report Model/Part/Revision
SS
Get Met One serial number
Table 6-1 Common Serial Command Set