3.5.10 Nozzle/Count Test
The Nozzle/Count Test screen provides the ability
to manually raise and lower the nozzle. Pressing
the grey NOZZLE button will cause the nozzle to
change state. When the nozzle is in the down
position, the E-BAM beta count results can also
be verified as shown in Figure 3-21. The
frequency for beta counts should be greater than
500 Hz.
Figure 3-21 The Nozzle/Count Test Screen
Counting should only occur when the nozzle is in the down position. Figure 3-21 shows a typical
count value for the Beta Count field.
The 1-Minute Count test is for troubleshooting purposes with factory support technicians and will not
commonly be used when deployed. Pressing the grey START button will begin the 1-Minute Count
test. The counts displayed in the 1-Minute Count Field will start from zero and begin counting up. A
60 second countdown timer will be shown in parenthesis next to the count value.
3.5.11 Digital Link
Test digital communications with the 597A by
entering this screen. The 597A firmware version
should be displayed when the proper link is
established. See section 2.7 for details on how
to configure the 597A sensor, as well as any
optional wind sensors that may be.