3.6 Setup Menu
The Setup Menu grants access to the configuration of most of the operating parameters for the E-
BAM. It allows for changing the offsets, clearing the memory, setting the date and time, and much
more. Because of the changes able to be made, the Setup Menu is password protected.
Figure 3-23 Setup Menus
Note that the display has a limited amount of space and cannot show all options on one screen. Tap
the down or up keys in the lower right corner of the display to navigate between the three screens
shown in Figure 3-23.
3.6.1 Set Clock
This is where the date and time are set. Press
the green box of the field that needs to be
modified. The numerical entry keypad (Figure
3-5) will be displayed and allow entry of the
value for that parameter. Once all fields have
been entered, press the grey Set button in the
lower left corner to set them.
Figure 3-24 The Set Clock Screen
3.6.2 Sample
The concentration (Conc) Units of the E-BAM
can be set to either mg/m
3
or μg/m
3
. To change
this setting, press the green bordered value box
and a list of the settings available will be
displayed. Select the desired units and the press
the grey OK button.
Figure 3-25 The Sample Screen
The Real Time Period sets the time interval for the real-time concentration (Conc RT) algorithm.
The Conc RT will use of the last X number of minutes where X is Real Time Period value set in the
Sample menu screen.
For example, if 15 minutes is chosen, nothing will be displayed for minutes 1-14. After minute 15,
the first Conc RT will be displayed. Each minute after that will update the Conc RT based on the
previous 15 minutes. This means that minute 16 will display the data from minute 2-16, and so on.
The Real Time Period may be set anywhere from 15 to 60 minutes, in one-minute increments.