47
Advanced functions
Setting:
g
Overtemperature limit:
up to max. 10 °C above nominal temperature
(for details of the nominal temperature, see nameplate)
Setting:
1. Select the
MAX
icon with the push-turn control.
2. Hold down the SET key and set the desired temperature limit
with the push-turn control (e.g.
38.5
°C).
°C
MIN
AUTO
MAX
Undertemperature limit:
Adjustment range: 10 °C below the minimum temperature of the
appliance up to 10 °C above nominal temperature of the appli-
ance (for details of the nominal temperature, see nameplate).
The lower alarm limit value cannot be set higher than the top
one. If no undertemperature limit is required, set the lowest
temperature.
°C
MIN
AUTO
MAX