30
Operation and control
5.10.2
Settings example normal mode
With a 5% CO
2
content, a 3% oxygen content and a humidity of 96%, the appliance should
heat up to 37°C. The monitoring function should respond at 38.5 °C ( Fig. 20 ).
30
40
20
10
50
60
80
40
20
100
T
e
mperature in °C
Humidity rh in %
Time
Time
Time
CO
2
in %
5
10
15
20
25
Time
O
2
in %
5
10
15
20
25
Monitoring temperature
Fig. 20
Example for normal mode
1. Setting the normal operating mode:
Hold SET key down for approx. 3 seconds, the current
operating mode then begins to flash.
Select the
operating mode with the push-turn con-
trol, while the SET key is held down. After you let go of
the SET key, the control is in the normal operating mode.
2. Setting the temperature setpoint:
Hold down the SET key and set the desired temperature
setpoint of
37.0
°C with the push-turn control.
Release the SET key, the appliance will briefly flash, show-
ing the temperature setpoint. Then the current tempera-
ture
appears on the display and the controller begins to move
to the set temperature of 37.0 °C.
Ź
Heating up is indicated by the
icon.
°C
3. Setting the monitoring temperature:
Turn the push-turn control to the right until the monitor-
ing temperature and the MIN or MAX icon flashes. Hold
down the SET key and with the push-turn control, set
the overtemperature limit to
38.5
°C and the undertem-
perature limit to
36.0
°C. Turn the push-turn control to
the right until the monitoring temperature and the AUTO
icons flash. Hold down the SET key and set to
on
with the
push-turn control.
The tolerance band is set in the SETUP menu (see
page 44).
°C
MIN
AUTO
MAX
SETUP