Operation and control
5.10.1 Normal mode
The appliance runs in this operating mode in permanent operation. The desired setpoints for
operating the chamber can be selected. The settings have an immediate effect on the func-
tions of the appliance.
1. Load incubator (see page 25 ).
2. Switch on appliance. To do this, press the push-turn control on the operating panel so that
it comes out of the appliance (see Fig. 16 on page 26 ).
3. Select the normal operating mode
with the push-turn control:
SETUP
4. As described above, set the individual parameters with the push-turn control and the set
key:
Temperature setpoint
p
p
Adjustment range:
20°C
to
C
50°C
°C
Temperature monitoring
p
g
Adjustment range:
MIN MAX AUTO
(see also page 45)
°C
MIN
AUTO
MAX
Humidity setpoint
y
p
Adjustment range:
Ź
For incubators with basic
fittings:
88
to
97
%rh
Ź
For models with humidity
module:
40
to
97
%rh
rh
%
High levels of air humidity in the interior can only be achieved without condensation if the
interior is thoroughly heated. Please set the humidity only when the temperature in the
interior has been stable for at least 30 minutes.
CO
2
setpoint
p
Adjustment range:
0
bis
10
%
(or models with CO
2
or premium
modules 0 to 20
0
to
20
%)
CO
2
O
2
setpoint
p
(only for models with O
2
mod-
ule)
Adjustment range:
Off
,
1
to
20
%
t2
t1
on
off
Mo
Tu
We
Th
Fr
Sa
Su
o
2
= 3 . 4 %